WILKERSON INSTRUMENT CO., INC.

MM4408 DC Input Plug-In Mighty Module
Subtractor Transmitter Y=(A-B)

Wici Logo
Home Button Wilkerson Instrument Co.    800-234-1343 Wilkerson Instrument Co Blog PageWilkerson Instrument Co LinkedIn PageShopping Cart

MM4408

MM4408
DC Input
Plug-In
Mighty Module
Subtractor Transmitter
Y=(A-B)

$390.00

The following information must be entered to complete your order:
See "Specifications" below for Input and Output range limits.

Power:

Input:
Zero Scale
Full Scale
Units
Output: Zero Scale
Full Scale
Units
  Isolation

Quantity:

 

Don't Forget Your Accessories
DMP011 11-Pin Socket for Panel or DIN-Rail Mounting $25.00
Quantity:
Specifications
CLP1 Hold Down Assembly $20.00
Quantity:
Specifications
HKB-HK2D-11 Explosion Proof Housing With DMP011 Installed $330.00
Quantity:
Specifications
DR1 DIN-Rail, 35mm Symmetrical, 39 inches (1 meter) $25.00
Quantity:
Specifications
TRK48 PVC Snap-Track, 4 ft, for MP011 $30.00
Quantity:
Specifications
 
  • Features
  • Description
  • Specifications
  • Dimensions
  • Connections
  • Provides DC Output Proportional to Math Function
  • 1 V or 1 mA Minimum Input Span
  • Unlimited Choice of Input / Output Ranges (Within Specified Range Limits)
  • Choice of Power Options
  • 10 +Year Warranty

The math function module provides a DC output voltage or current proportional to the sum or difference of the DC inputs. Typical input ranges are 4/20 mA, 10/50 mA, 0-1 V, 0-5 V and 0- 10 VDC. Other ranges are available down to minimum spans of 1 V or 1 mA DC.

Standard calibration provides 0% output when all inputs are at 0%. Full scale (100%) output occurs when all additive inputs are at 100% and all subtractive inputs are at 0%. Other  calibrations are available.

All Wilkerson products are designed with RFI filters and lighting protection to reduce susceptibility to electrical noise and damage by lighting.

TYPICAL APPLICATIONS
Adding and subtracting flows, weights, etc. for process monitoring, control and data acquisition.

INPUT RANGE
Voltage select any range between ±250 V
(min span 1 V)
Typical inputs: 0-1 V, 0-5 V, 0-10 V
Current select any range between ±100 mA max
(min span 1 mA)
Typical inputs: 4/20 mA or 10/50 mA

INPUT IMPEDANCE
Voltage 100 kilohms
Current 1.25 V drop typical
others available

OUTPUT RANGE
Voltage select any range from -10 V to +15V
10 mA max load (min span 0.2 V)
Current select any range from 0 to 50 mA max
>24 V compliance (1200 ohms max at 20 mA)
18 V compliance for ISO option if full scale output >20 mA

RESPONSE TIME
(Range Dependent) < 100 ms

 

BALANCE BETWEEN INPUTS
< 0.2%

LINEARITY
±0.01% of span

ISOLATION
Breakdown, Power / Circuitry >1500 VAC rms

OPERATING TEMPERATURE
14°F to 140°F/–10°C to 60°C

TEMPERATURE STABILITY
±0.02% of span/°C max

POWER
115 VAC ±10%, 50/60 Hz (2.5 W max)
230 VAC ±10%, 50/60 Hz (2.5 W max)
(DC Power Option)
24 VDC (limits 21-32 VDC)
12 VDC (limits 10-16 VDC)
Isolation, DC power supply to input common: 10 megohms

Mighty Module Dimension Drawing

PIN 1
PIN 2
PIN 3
PIN 4
PIN 5
PIN 6
PIN 7
PIN 8
PIN 9
PIN 10
PIN 11
AC L1 or DC +
No Connection
AC L2 or DC -
Input A +
Input Common
Input B +
No Connection
No Connection
Output +
Output -
No Connection
Specifications    Manual

Solutions Provider for Signal Conditioning and
Process Control Instrumentation