WILKERSON INSTRUMENT CO., INC.

MM1200 Thermocouple Input Plug-In
Single Alarm Mighty Module

Wici Logo
Home Button Wilkerson Instrument Co.    800-234-1343 Wilkerson Instrument Co Blog PageWilkerson Instrument Co LinkedIn PageShopping Cart

MM1200

MM1200
Thermocouple Input
Plug-In
Single Alarm
Mighty Module

$365.00

The following information must be entered to complete your order:
See "Specifications" below for Input range limits.

Power:

Input:
Zero Scale
Full Scale
Units
  Thermocouple Type
Alarm: Alarm Action
  Alarm Logic

Quantity:

 

Don't Forget Your Accessories
DMP011 11-Pin Socket for Panel or DIN-Rail Mounting $25.00
Quantity:
Specifications
CLP1 Hold Down Assembly $20.00
Quantity:
Specifications
HKB-HK2D-11 Explosion Proof Housing With DMP011 Installed $330.00
Quantity:
Specifications
DR1 DIN-Rail, 35mm Symmetrical, 39 inches (1 meter) $25.00
Quantity:
Specifications
TRK48 PVC Snap-Track, 4 ft, for MP011 $30.00
Quantity:
Specifications
 
  • Features
  • Description
  • Specifications
  • Dimensions
  • Connections
  • Provides a DPDT Relay Contact Closure at a Preset Thermocouple Input
  • Cold Junction Compensated
  • Standard Fail-Safe Operation
  • Red and Green LED Alarm Status Indicators
  • Adjustable Deadband
  • Latching Alarm Available (MM1210)
  • 5 mV Minimum Input Span
  • Unlimited Choice of Input / Output Ranges ( Within specified range limits )
  • Upscale Burnout Protection (Downscale Available)
  • Choice of Power Options
  • 10 Year Warranty

The MM1200 monitors any thermocouple input signal and trips a dpdt, 5 A relay when the input exceeds the desired level. Normal operation has the relay energized for the non-alarm condition and de-energized for an alarm condition. This provides a fail-safe alarm condition for loss of power to the module. The alarm has a set of red/ green LEDs to indicate the alarm status.

A deadband adjustment allows a deadband of 0.5% to 100% of span to be set into the module. The deadband is symmetrical about the setpoint.

With the latching option, the alarm has no deadband control. Once the limit has been reached, the alarm latches and power must be momentarily interrupted to reset the alarm.

Cold junction compensation is provided by a solid state temperature sensor embedded in the thermocouple terminal strip.

All Wilkerson products are designed with RFI filters and lightning protection to reduce susceptibility to electrical noise and damage by lightning.

Upscale burnout protection is provided as standard. In the event the thermocouple opens, the module behaves as though the input goes offscale high. Option B provides downscale burnout
protection (module behaves as though the input goes low).

TYPICAL APPLICATIONS
Heater/cooler control, HI/LO temperature alarm.

INPUT RANGE
select any type thermocouple (min span 5 mV)

SETPOINT
0 to 100% of span

DEADBAND
0.5% to 100% of span

RELAY CONTACTS (dpdt)
Resistive Load 5 A max,
   150 W max, 220 VAC max, 30 VDC max
Inductive Load (power factor >0.4)
   2.5 A max, 75 W max, 220 VAC max, 30 VDC max

ACCURACY
±0.1% of span

TRANSISTOR OUTPUT (Option V)
relay driver (12 V coil, >220 ohms) or open-collector outputs
sink 100 mA, 30 V supply max

COMMON MODE REJECTION
120 dB, DC to 60 Hz

OPERATING TEMPERATURE
14°F to 140°F/–10°C to 60°C

TEMPERATURE STABILITY
±(0.02% of span + 1.3 μV)/°C max

POWER
115 VAC ±10%, 50/60 Hz (2.5 W max)
230 VAC ±10%, 50/60 Hz (2.5 W max)
(DC Power Option)
24 VDC (limits 21-32 VDC)
12 VDC (limits 10-16 VDC)

Isolation, DC power supply to input common: 10 megohms

 

Thermocouple Input Mighty Module Dimension Drawing

T/C Input +
T/C Input +
PIN 1
PIN 2
PIN 3
PIN 4
PIN 5
PIN 6
PIN 7
PIN 8
PIN 9
PIN 10
PIN 11
T/C Terminal+
T/C Terminal -
Power AC L1 or DC +
No Connection
Power AC L2 or DC -
No Connection
No Connection
Relay Set 1 NO
Relay Set 1 C
Relay Set 1 NC
Relay Set 2 NO
Relay Set 2 C
Relay Set 2 NC
Specifications    Manual

Solutions Provider for Signal Conditioning and